Lambdometer detector

L. S. Vorob'ev, V. B. Gavrilov, N. A. Goryainov, Yu G. Grishuk, P. V. Degtyarenko, Yu V. Efremenko, B. V. Zagreev, M. V. Kosov, S. G. Kuznetsov, S. V. Kuleshov, G. A. Lekcin, N. A. Pivnyuk, A. V. Smirnitskij, V. B. Fedorov, F. M. Khasanov, B. B. Shartsman, S. M. Shuvalov

Research output: Contribution to journalArticlepeer-review

1 Citation (Scopus)

Abstract

The design and characteristics of 'Lambdometer' detector for investigation of cumulative lambda-hyperons in the range of fragmentation of resting target nucleus in hadron-nuclear interactions under various initial conditions were described (primary particle energy and kind, target nucleus atomic number, lambda-particle escape angle). The detector permitted to measure lambda-particle spectra in the kinetic energy range from 40 to 150 MeV with the resolution of approximately 5 MeV and relative sistematic error of <25%. It made it possible to determine lambda-particle cross-section (yield) with the same accuracy and polarization with the accuracy of ≥96%.
Original languageEnglish
Pages (from-to)49-58
Number of pages10
JournalPribory i Tekhnika Eksperimenta
Publication statusPublished - 1 Mar 1991
Externally publishedYes

ASJC Scopus subject areas

  • Nuclear and High Energy Physics

Fingerprint

Dive into the research topics of 'Lambdometer detector'. Together they form a unique fingerprint.